Edge launch SMAs can short into the internal copper layers with some installation methods. Watch for these gotchas.
Category Archives: Flagship DL01
Flagship DL01: Continuous Acquisition with Overridden Tx Gain
As of now 7/5/24 with customer interface software with the DL0100A1, you can stream data in continuous acquisition mode with the higher (max override) transmit gains (Tx gains aka pulse voltages) using the following steps:
Flagship DL01: “Damping” Section: Values and Selections
You could call this part of the board by various names, most of which are printable. Here, we’re just calling it “damping”. It’s rudimentary and is there to be adjusted across assembly variants and system needs, with some basic runtime operability. It can tune the output and cable/XD interface a bit. Mostly, from early work, …
Continue reading “Flagship DL01: “Damping” Section: Values and Selections”
Flagship DL01: Pulse Voltage Table
Tx Gain Setting DL-200V DL-300V DL-500V 0 42 64 103 1 57 85 141 2 71 107 180 3 88 131 224 4 123 182 320 5 147 218 387 6 219 323 585 7 270 408 (!) Gain setting to “tank” voltage table. Please see notes below. Important details follow – please click read …
Flagship DL01: Pulse Voltage
This post will indicate how you can measure the voltage available for pulsing the transducer for different settings. BUT, there are important considerations in understanding what the transducer actually experiences.